The quality of sample extraction and preparation significantly impact MS results. For gel bands, please carefully follow the following instructions:
CRITICAL: Special care must be taken to avoid contamination in every step, especially with keratins from skin or hair (always wears clean nitrile gloves and work in a dust- free environment).
CRITICAL: DO NOT use silverstaining! Only Coomassie-stained gels!
CRITICAL: PRECISELY cut out ONLY the band of interest (only the stained area) - any excess gel will lead to background noise. If the band is extremely faint, there is a likely chance that we won't get any protein identification. So, we suggest you load the maximum possible amount of protein sample into the gel so that we get a fairly visible band with Coomassie stain.
Email us a gel picture before sending the samples. This will help us to understand whether the band is properly stained or not. This could also be used as a reference to understand the sample concentration and its complexity. Since the nano-LC is a highly sensitive instrument, it is very important not to overload the nano-LC column with excess sample for the optimal results.
ONLY cut gels on a clean glass plate (never use overhead projector foils or aluminium foils). The glass plate can be cleaned by using organic solvents like 70% ethanol or methanol.
Use a NEW clean scalpel blade for precise cutting of gel spots.
Always use filtered deionized water.
DO NOT wash any flasks, tubes, or glass plates for electrophoresis with soap (or any polymeric detergent). Always, rinse your glassware with hot water and then an organic solvent like 70% ethanol or methanol.
Freshly prepare all the buffers and stains needed to run the gel for MS analysis (DO NOT re-use).
Place the gel cubes in a clean microcentrifuge tube (1.5 ml) and add enough methanol (50% methanol) in it to cover the gel cubes. The addition of large volumes of the buffer is not required, having them moist is enough.
Make sure that the lid of the microcentrifuge tube is CLOSED PROPERLY before placing it in an envelope to send.
DO NOT use parafilm around the lids.
You can send the samples by courier or ordinary post. There is no need to ship the samples on ice. The samples are stable in 50% methanol.